Summary of Contents for National Instruments 776572-1600
Page 1
SCXI SCXI-1125 User Manual SCXI-1125 User Manual October 2009 372425C-01...
Page 2
Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545 For further support information, refer to the Technical Support Information document. To comment on National Instruments documentation, refer to the National Instruments Web site at and enter the ni.com/info...
Page 3
Instruments Corporation. National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.
Page 4
Conventions The following conventions are used in this manual: <> Angle brackets that contain numbers separated by an ellipsis represent a range of values associated with a bit or signal name—for example, P0.<3..0>. Square brackets enclose optional items—for example, [ response »...
Chapter 1 About the SCXI-1125 What You Need to Get Started ..................1-1 National Instruments Documentation ................1-3 Installing Application Software, NI-DAQ, and the DAQ Device .........1-4 Installing the SCXI-1125 Module into the SCXI Chassis.......1-4 Connecting the SCXI-1125 in an SCXI Chassis to an E/M/X Series DAQ Device for Multiplexed Scanning ............1-4...
Page 6
Contents Chapter 3 Configuring and Testing SCXI-1125 Software-Configurable Settings ..............3-1 Common Software-Configurable Settings ............3-1 Filter Bandwidth ................3-1 Gain/Input Range................3-1 Connecting the SCXI-1125 in an SCXI Chassis to an E/M/X Series DAQ Device for Multiplexed Scanning............3-2 Connecting the SCXI-1125 in a PXI/SCXI Combination Chassis to an E/M/X Series DAQ Device for Multiplexed Scanning........
Page 8
Contents Appendix A Specifications Appendix B Using SCXI Channel Strings with Traditional NI-DAQ (Legacy) 7.0 or Later Appendix C Removing the SCXI-1125 Appendix D Common Questions Glossary Index Figures Figure 2-1. Connecting a Ground-Referenced Signal ..........2-2 Figure 2-2. Connecting a Floating Signal..............2-3 Figure 2-3.
Page 11
Chapter 1 About the SCXI-1125 For maximum allowable voltage for a particular terminal block refer to Table A-2, Note Terminal Block Maximum Voltages. – SCXI or PXI/SCXI combination chassis – One of the following: • E/M/X Series multifunction DAQ device •...
• Software documentation—You may have both application software and NI-DAQ software documentation. National Instruments (NI) application software includes LabVIEW, LabWindows/CVI, and Measurement Studio. After you set up the hardware system, use either your application software documentation or the NI-DAQ documentation to help you write your application.
Chapter 1 About the SCXI-1125 • One or more of the following help files for software information: – Start»All Programs»National Instruments»NI-DAQ» NI-DAQmx Help – Start»All Programs»National Instruments»NI-DAQ» Traditional NI-DAQ User Manual – Start»All Programs»National Instruments»NI-DAQ» Traditional NI-DAQ Function Reference Help •...
Chapter 1 About the SCXI-1125 If you have already installed the appropriate software, you are ready to configure the SCXI-1125 module(s) you installed for parallel mode operation. Verifying the SCXI-1125 Installation in Software Refer to the SCXI Quick Start Guide for information about verifying the SCXI installation.
Page 17
Chapter 1 About the SCXI-1125 If the cabled module is not listed in the correct slot, complete the following troubleshooting steps: If the cabled module is not listed in the correct slot and the slot is empty, click the drop-down listbox next to the correct slot and select the cabled module.
Chapter 2 Connecting Signals connections to each channel are fully floating with respect to ground and completely isolated from other channels. You can operate with common-mode voltage levels up to 300 V Figures 2-1 through 2-4 show signal connection methods that give the highest noise immunity.
Chapter 2 Connecting Signals AC-Coupling You can have an application where you wish to measure only AC voltages and remove the DC component of a signal before amplification and sampling. In such cases, you can connect a capacitor in series with one or both input terminals of the SCXI-1125, as shown in Figures 2-3 and 2-4.
Chapter 2 Connecting Signals Table 2-1. Front Signal Pin Assignments Front Connector Diagram Pin Number Column A Column B Column C AI 0 + — AI 0 – — — — Column AI 1 + — AI 1 – — —...
Chapter 3 Configuring and Testing Connecting the SCXI-1125 in an SCXI Chassis to an E/M/X Series DAQ Device for Multiplexed Scanning Refer to the SCXI Quick Start Guide to install the cable adapter and connect the SCXI modules to the DAQ device. If you have already installed the appropriate software, refer to Chapter 3, Configuring and Testing, to configure the SCXI-1125 module(s).
Chapter 3 Configuring and Testing Select Analog Input. Select Voltage. If you are creating a task, you can select a range of channels by holding down the <Shift> key while selecting the channels. You can select multiple individual channels by holding down the <Ctrl> key while selecting channels.
Chapter 3 Configuring and Testing Creating a Virtual Channel To create a virtual channel, complete the following steps: Right-click Data Neighborhood and select Create New. Select Traditional NI-DAQ Virtual Channel and click Finish. Click Add Channel. Select Analog Input from the drop-down list and click Next. Enter the Channel Name and Channel Description, and click Next.
Chapter 3 Configuring and Testing Use the format n to verify the signal, where n is a ob x ! sc y ! md z ! single input channel. Complete the following steps to use channel strings in verifying the signal: Expand Devices and Interfaces.
SCXI-1125. Using this mode of operation, you can scan all input channels of the SCXI-1125 into one output channel that is read by the National Instruments DAQ device. You can also multiplex the CJC channel that connects to a sensor on the SCXI terminal block for making temperature measurements.
Chapter 4 Theory of Operation Operating in Parallel Mode You can configure the SCXI-1125 to operate in parallel mode as described in Chapter 1, About the SCXI-1125. In parallel mode, all eight analog output channels on the SCXI-1125 are connected to eight analog input channels on the E/M/X Series DAQ device.
Page 39
You have completed the steps to get the true temperature reading from the thermocouple junction. National Instruments software ADEs have useful conversion functions for CJC. In LabVIEW, virtual channels with the CJC channel invoked or the Convert Thermocouple Reading VI are used. In C, use the NI-DAQ function, .
Chapter 5 Using the SCXI-1125 Making High-Voltage Measurements Another common use of the SCXI-1125 is to make measurements up to 1000 VDC. Making measurements beyond ± 5 V requires use of the an attenuator terminal block. The SCXI-1327 and SCXI-1313A terminal blocks have a selectable attenuator for choosing between no attenuation or 100:1 attenuation, which allows you to use the SCXI-1125 with up to 300 V...
Chapter 5 Using the SCXI-1125 Create Task Using DAQ Assistant? Create a Task Programmatically Create Task in DAQ Assistant or MAX Create Channel Create Another (Application Specific) Channel? Hardware Timing/Triggering? Further Configure Channels? Adjust Timing Settings Configure Channels Analyze Data? Process Start Measurement Data...
Chapter 5 Using the SCXI-1125 When programmatically adjusting timing settings, you can set the task to acquire continuously, acquire a buffer of samples, or acquire one point at a time. For continuous acquisition, you must use a while loop around the acquisition components even if you configured the task for continuous acquisition using MAX or the DAQ Assistant.
Chapter 5 Using the SCXI-1125 Table 5-4. Programming a Task in LabVIEW Flowchart Step VI or Program Step Create Task in DAQ Assistant Create a located on the DAQmx Task Name Constant Controls»All Controls»I/O»DAQmx Name Controls subpalette, right-click it, and select New Task (DAQ Assistant). Create a Task —This VI is optional if you created DAQmx Create Task.vi...
Chapter 5 Using the SCXI-1125 Change the property to read or write to either get the property or write a new value. Right-click the property, go to Change To, and select Write, Read, or Default Value. After you have added the property to the property node, right-click the terminal to change the attributes of the property, add a control, constant, or indicator.
Chapter 5 Using the SCXI-1125 Text Based ADEs You can use text based ADEs such as LabWindows/CVI, Measurement Studio, Visual Basic, .NET, and C# to create code for using the SCXI-1125. LabWindows/CVI LabWindows/CVI works with the DAQ Assistant in MAX to generate code for a task.
Chapter 5 Using the SCXI-1125 Traditional NI-DAQ (Legacy) in LabVIEW LabVIEW is a graphical programming environment for test and measurement application development with built-in easy to use tools for data acquisition, analysis, and display. You can use functional graphical blocks called subVIs to easily create a custom application that fully utilizes the SCXI-1125 programmable functionality.
Chapter 5 Using the SCXI-1125 Configure the SCXI-1125 Settings Using Traditional NI-DAQ (Legacy) in LabVIEW You can configure SCXI-1125 settings in MAX using the virtual channel. To configure and control the SCXI-1125 from LabVIEW, use the AI Parameter VI. You can find AI Parameter VI in the function subpalette Data Acquisition»Analog Input»Advanced Analog Input.
Chapter 5 Using the SCXI-1125 Convert Scaling Using Traditional NI-DAQ (Legacy) in LabVIEW If you need scaling, you can either use an analog input voltage virtual channel with a custom scale configured in MAX or SCXI channel strings, and provide scaling in your LabVIEW application. If you are using SCXI channel strings, you can easily convert the SCXI-1125 voltage signal measurements in your application into scaled units of interest such as pounds or newtons.
NI-DAQ function calls shown in Table 5-8. For more information regarding each setting, refer to the Traditional NI-DAQ (Legacy) Function Reference Help installed by default in Start»All Programs»National Instruments»NI-DAQ. SCXI-1125 User Manual 5-22 ni.com...
For more information regarding each acquisition function, refer to the Traditional NI-DAQ (Legacy) Function Reference Help installed by default in Start»All Programs»National Instruments» NI-DAQ. Perform Scaling, Analysis, and Display...
Chapter 5 Using the SCXI-1125 The last parameter, channels, is the list of channels that are scanned for module z. It can have several formats: • n, where n is a single input channel. ob x ! sc y ! md z ! •...
Chapter 5 Using the SCXI-1125 C and Low-Level DAQ Functions When using a C-based environment, several steps are needed to configure the SCXI-1125 for multiplexed scanning. The following procedure outlines the steps for programming with the low-level DAQ function calls: Prepare the SCXI-1125 settings by either loading the original SCXI configuration settings using , or by specifying...
By default, the example programs are installed in C:\Program . More examples Files\National Instruments\CVI x . x \Samples are installed by default in C:\Program Files\National Instruments\NI-DAQ\Examples Traditional NI-DAQ (Legacy) Measurement Studio Examples Many example programs ship with NI-DAQ.
SCXI-1125 module. Calibration Procedures You can calibrate the offset on the SCXI-1125 using National Instruments software. When calibrating the offset on the SCXI-1125, make sure the DAQ device you are using has been calibrated recently or you will invalidate the offset calibration on the SCXI-1125.
Chapter 5 Using the SCXI-1125 Table 5-9. Gain Values and Input Limits (Continued) Gain Range (V) 1000 ±0.005 2000 ±0.0025 If you are using a C-based ADE, use the following procedure to do an offset calibration on the SCXI-1125: Make sure the DAQ device or DMM you are using has a valid calibration and meets the accuracy specifications for your application.
Page 71
Appendix A Specifications Overvoltage protection Isolated connector pins: Powered on and off......±300 V Inputs protected .......CH0..CH7 Non-isolated connector pins: Powered on and off......+5.5V/–0.5 V SCXI-1125 User Manual ni.com...
Page 75
Appendix A Specifications Common mode rejection ratio, 50/60 Hz (CMRR) 4 Hz filter enabled ......160 dB 10kHz filter enabled ......98 dB Normal mode rejection, 50/60 Hz (NMRR) 4 Hz filter enabled ......60 dB Crosstalk at 1kHz Adjacent channels......–75 dB All other channels......–90 dB Input coupling Default ..........DC Using SCXI-1305 or TBX-1329 ..AC or DC...
Appendix A Specifications Maximum Working Voltage Maximum voltage rating refers to the signal voltage plus the common mode voltage (Signal + common mode). Voltage of each input shall remain within 300 V of ground. Channel-to-earth ........300 V, Measurement Category II Channel-to-channel.........300 V, Measurement Category II Table A-2.
Page 79
EU Customers recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit ni.com/environment/weee National Instruments (RoHS) National Instruments RoHS ni.com/environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china...
Page 81
Appendix B Using SCXI Channel Strings with Traditional NI-DAQ (Legacy) 7.0 or Later is the list of channels that are scanned for module . It can have channels several formats: • , where is a single input channel. ob x ! sc y ! md z ! nx •...
Page 83
Appendix C Removing the SCXI-1125 If the SCXI-1125 is the module cabled to the E/M/X Series DAQ device, disconnect the cable. Remove any terminal block that connects to the SCXI-1125. Rotate the thumbscrews that secure the SCXI-1125 to the chassis counterclockwise until they are loose, but do not completely remove the thumbscrews.
Page 85
Appendix D Common Questions Which version of NI-DAQ is needed to work with the SCXI-1125 and how do I get the most current version of NI-DAQ? You must have NI-DAQmx 9.0 or later. Visit and follow the link, ni.com Download Software»Drivers and Updates»Search Drivers and Updates, and type in the keyword to find the latest version of NI-DAQ...
Page 87
Appendix D Common Questions In LabVIEW, can I use a VI to change my filter setting? In NI-DAQmx, you can change the filter settings using a DAQmx Channel property node. In Traditional NI-DAQ (Legacy), there is no VI available to do this.
Page 89
Glossary analog-to-digital absolute accuracy The maximum difference between the measured value from a data acquisition device and the true voltage applied to the input, typically specified as ±voltage. alternating current analog-to-digital converter—An electronic device, often an integrated circuit, that converts an analog voltage to a digital number. application development environment—A software environment incorporating the development, debug, and analysis tools for software development.
Page 91
Glossary Data acquisition—(1) collecting and measuring electrical signals from sensors, transducers, and test probes or fixtures and inputting them to a computer for processing; (2) collecting and measuring the same kinds of electrical signals with A/D and/or DIO boards plugged into a computer, and possibly generating control signals with D/A and/or DIO boards in the same computer.
Page 93
Glossary isolation A type of signal conditioning in which you isolate the transducer signals from the computer for safety purposes. Isolating the signals protects you and your computer from large voltage spikes and makes sure the measurements from the DAQ device are not affected by differences in ground potentials.
Page 94
NI-DAQ The driver software needed to use National Instruments DAQ devices and SCXI components. noise Analog. Unwanted signals. Noise comes from both external sources, such...
Page 95
A rugged, open system for modular instrumentation based on CompactPCI, with special mechanical, electrical, and software features. The PXIbus standard was originally developed by National Instruments in 1997, and is now managed by the PXIbus Systems Alliance. resolution The smallest signal increment that can be detected by a measurement system.
Page 97
Glossary thermocouple A temperature sensor created by joining two dissimilar metals. The junction produces a small voltage as a function of the temperature. typical Underwriters Laboratory volts volts direct current virtual instrument—(1) a combination of hardware and/or software elements, typically used with a computer, that has the functionality of a classic standalone instrument (2) a LabVIEW software module (VI), which consists of a front panel user interface and a block diagram program.
Page 99
Index DC voltage connections. See AC and DC voltage connections input characteristics, A-1 digital lines, unavailability on DAQ installation device, D-2 connecting SCXI-1125 to DAQ device for digital signals on SCXI-1125 (table), D-3 parallel scanning, 1-5 documentation connecting to DAQ device for conventions used in the manual, iv multiplexed scanning in PXI combination chassis, 1-5, 3-2...
Page 101
Index software multiplexed scanning operations, 5-24 temperature measurements using LabVIEW and SCXI channel thermocouples, 5-1 string, 5-25 accurate method for temperature LabVIEW and virtual channel determination, 5-2 string, 5-26 guide for calculating overall temperature parallel scanning operations error, 5-3 C and parallel mode, 5-29 overview, 5-1 LabVIEW and parallel mode, 5-28 temperature sensor connection, 2-7...