Watch Dog - Siemens LDS 6 Operating Instructions Manual

In situ laser gas analyzers
Table of Contents

Advertisement

Operation

4.4 Watch dog

See also
Maintenance request alarm (Page 60)
4.4
Watch dog
In the unlikely event that the software in LDS 6 should crash it will reboot with the help of a watch
dog. The instrument will be down for approximately 3 minutes if this should happen and then
measure normally again. It is very unusual that this should happen but if it does the watchdog
prevents the system from going down and stay down.
Module crash
Measurement
OK
Figure 4-4
Illustration of the events during a software crash.
58
From this point, normal
startup behaviour
Frozen value
0 mA output
0-16 s
approx 1 min
Function control
mode (configurable)
0 mA out if response
is latest or if Function
control is off
approx 1 min
Operating Instructions, 01/2022, A5E00295894-10
Measurement
OK
LDS 6

Advertisement

Table of Contents
loading