2. SPECIFICATIONS
Range
Resolution
Titrator
Accuracy
Sample size
Titration method
Principle
Pump speed
Stirring speed
Log data
Range
Resolution
Accuracy
pH Meter
Calibration
Temperature
compensation
Range
Resolution
mV Meter
Accuracy
Log data
Low Range
%I.a.: 0.01 to 0.20
°SH: 0.4 to 8.9
°D: 1.0 to 20.0
°TH: 1.1 to 22.2
High Range
%I.a.: 0.1 to 2.0
°SH: 4.4 to 88.9
°D: 10 to 200
°TH: 11.1 to 222.2
%I.a.: 0.01 (Low Range) / 0.1 (High Range)
°SH: 0.1
°D: 0.1 (Low Range) / 1 (High Range)
°TH: 0.1
Low Range
± 0.01 %l.a. @ 25 °C
High Range
± 0.1 %l.a. @ 25 °C
Low Range 20: 20 mL or 20 g
Low Range 50: 50 mL or 50 g
High Range 20: 20 mL or 20 g
Acid-base titration
Endpoint titration
10 mL/min
800 (Low Range) / 1000 (High Range)
Up to 200 samples
-2.0 to 16.0 pH / -2.00 to 16.00 pH
0.1 pH / 0.01 pH
± 0.01 pH
1, 2 or 3 point calibration;
4 available buffers (4.01, 6.00, 8.30, 10.01)
Manual or automatic
-2000.0 to 2000.0 mV
0.1 mV
± 1.0 mV
Up to 200 samples (pH or mV)
5