SPECIFICATIONS

Titrator
Range
Resolution LR: 0.1 mg/L / 0.1 meq/L
Accuracy LR: ± 1 mg/L or 3% of reading @ 25 °C whichever is greater
Sample Volume 50 mL
Titration method Acid-base titration
Principle
Pump speed
Stirring speed
Log data
pH meter
pH meter
pH resolution:
pH accuracy:
pH calibration: 1, 2 or 3 calibration points;
Temperature
Compensation
mV meter
mV meter
mV resolution: 0.1 mV
mV accuracy:
Log data
Temperature
Range
Resolution
Accuracy
HI 1131B
Electrode
Temperature Probe HI 7662-T
Environment
0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply
12 Vdc power adapter
Dimensions
235 x 200 x 150 mm (9.2 x 7.9 x 5.9")
Weight
1.9 kg (67.0 oz.)
Low Range (LR): 30.0 - 400.0 mg/L CaCO
0.6 - 8.0 meq/L CaCO
High Range (HR): 300 - 4000 mg/L CaCO
6.0 - 80.0 meq/L CaCO
HR: 1 mg/L / 0.1 meq/L
HR: ± 10 mg/L or 3% of reading @ 25 °C whichever is greater
End point titration pH 4.50 or pH 8.30
10 mL/min
600 rpm
Up to 200 samples
-2.0 to 16.0 pH / -2.00 to 16.00 pH
0.1 pH / 0.01 pH
± 0.01 pH
4 available buffers (4.01, 7.01, 8.30, 10.01)
Manual or automatic
-2000.0 to 2000.0 mV
± 1.0 mV
Up to 200 samples (pH or mV)
-20.0 to 120.0 °C (-4.0 to 248.0 °F)
0.1 °C
±0.4 °C without probe error
6
3
3
3
3