7. SETUP MENU
The titrator's setup menu may be accessed from the main screen (meter or titrator) by pressing the
MENU key, then Setup.
A list of setup parameters will be displayed with currently configured setting.
While in the Setup menu, it is possible to modify the instrument's operation parameters. The ARROW
keys permit the user to scroll the setup parameters.
Press HELP to view the contextual help.
Press ESC to return to the main screen.
Concentration Unit
% l.a., °SH, °D or °TH
Press the corresponding virtual option key to change the option.
% l.a. - % lactic acid
°SH - °Soxhlet Henkel
°D
- °Dornic
°TH - °Thörner
Acidity Titration
Low 20, Low 50 or High 20
Press the corresponding virtual option key to select the desired range.
Note: Different titrants are required for each range.
Sample Type
Mass or Volume
Samples can be added by volume or by mass. For improved accuracy all samples should be added
by mass.
14