Specifications; Precision And Accuracy - Hanna Instruments HI 96704C Hydrazine ISM Instruction Manual

Hide thumbs Also See for HI 96704C Hydrazine ISM:
Table of Contents

Advertisement

www.ponpe.com
0 to 400 µg/L
Range
1 µg/L
Resolution
Accuracy
±3% of full scale
Typical EMC Deviation ±2 µg/L
Light Source
Light Emitting Diode
Light Detector
Silicon Photocell with narrow band interference filter @ 466nm
Method
Adaptation of the ASTM Manual of Water and Environmental Technology,
method D1385-88 for natural and treated water. The reaction between
hydrazine and the liquid reagent causes a yellow tint in the sample.
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type
1 x 9 volt
After 10' of non-use in measurement mode ;
Auto-Shut off
after 1 hour of non-use in calibration mode ;
with last reading reminder.
Dimensions
192 x 102 x 67 mm (7.6 x 4 x 2.6")
Weight
290 g (10 oz.).
REQUIRED REAGENTS
Code
HI 93704-0

PRECISION AND ACCURACY

Precision
Precision
Precision
Precision is how closely repeated measurements
Precision
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy
Accuracy
Accuracy is defined as the nearness of a test
Accuracy
Accuracy
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
100 µg/L hydrazine and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 3 µg/L.

SPECIFICATIONS

Description
Liquid Reagent
5
Quantity/test
2x12 drops