Specifications; Precision And Accuracy - Hanna Instruments HI 96713C Instruction Manual

Phosphate low range ism
Hide thumbs Also See for HI 96713C:
Table of Contents

Advertisement

www.ponpe.com
Range
0.00 to 2.50 mg/L
Resolution
0.01 mg/L
±0.04 mg/L ± 4% of reading @ 25°C
Accuracy
Typical EMC Deviation ±0.01 mg/L
Tungsten lamp
Light Source
Light Detector
Silicon Photocell with narrow band interference filter @ 610 nm
Method
Adaptation of the ascorbic acid method. The reaction between
phosphate and the reagent causes a blue tint in the sample.
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type
1 x 9 volt
After 10' of non-use in measurement mode ;
Auto-Shut off
after 1 hour of non-use in calibration mode ;
with last reading reminder.
Dimensions
192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
Weight
360 g (12.7 oz.).
REQUIRED REAGENTS
Code
HI 93713-0

PRECISION AND ACCURACY

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
1.00 mg/L phosphate and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.03 mg/L.

SPECIFICATIONS

Description
Powder Reagent
5
Quantity/test
1 packet
contact:
sale@ponpe.com