8.3. TITRATION MESSAGES

This screen is displayed when the instrument is in titration
mode. Press Start to begin a titration, Meter to enter pH
meter mode or Prime to enter into the prime function.
The titration result, expressed as concentration in selected
unit, is displayed automatically at the end of the titration.
Press Restart to start another titration or ESC to return to
the main screen.
This error message appears when the input reading (pH
or temperature) exceeds the specified limits. The pH or
temperature value and concentration will blink indicating
an error.
This screen appears when the sample concentration is out
of range.
This screen appears when the stirrer is not working properly.
Check the stir bar and beaker content. Press Restart to try
again.
This error message appears when the pump is not working
properly. Check the tubing, valve and syringe. Press Restart
to try again.