Ph Meter Glp Information - Hanna Instruments HI84529 Instruction Manual

Titratable acidity mini titrator & ph meter for dairy products
Table of Contents

Advertisement

Use ARROW keys when is displayed to scroll between the records.
Press ESC to return to the previous screen.
If Delete or Del.All is pressed, the instrument will ask for confirmation.
Press Yes to delete the record or No to return to the previous screen without deleting.
Deleting a single record will renumber the list of records.
If the pH log is empty, the message "No records available!" will be displayed.

16.2. pH METER GLP INFORMATION

The pH meter GLP screen displays the pH calibration data.
To view this information press MENU key while in pH meter mode then GLP.
Press Electrode to view information regarding electrode calibration.
The following items are included in electrode GLP: the time and date of the last calibration, offset,
slope, electrode condition, calibration timeout and the calibration buffers. The buffers displayed in
video inverse mode are from the previous calibration.
If a calibration hasn't been performed, the message "Not Calibrated" will be displayed.
44

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hi84529u-01

Table of Contents