Installation guidelines

4.1

Installation

This chapter provides instructions for the installation of the LDS 6, Optics Housing CD 6 and its
cables. The figure below shows the principle set-up of the LDS 6 central unit, optics housings and
cables.
Figure 4-1
Note
Do not install the optics housings unless the purging media is accessible (for optics housings that
require purging). Never leave a optics housing without purging since the wedge optics is easily
damaged.
At installation of the cross duct optics housing, the fiber end and the photo-detector have to be
aligned to the optical axis of the optics housing pair. Note that each optics housing has an optical
axis of its own which is its axis of symmetry. Normally the process flanges are mounted with an
angle error. The optics housings are however, equipped with a flange pair with a spherical
interface. This is used when the optics housing pair is aligned to each other. In this way an angle
error of up to ±2° on each side can be adjusted for.
Installation of optics housings
1. Check that the flanges are installed properly, according to the instructions earlier in this
manual.
2. Mount the optics housings, receiver and transmitter, with gaskets on the flanges and cross
tighten the bolts.
See also
Alignment (Page 20)
LDS 6, Optics Housing CD 6
Operating Instructions, 01/2022, A5E01134377-05
Principle set-up of an installation