Technical description

5.3 LDS 6, Optics Housing CD 6

5.3
LDS 6, Optics Housing CD 6
The LDS 6, Optics Housing CD 6 (shown in the figure below) is designed for in situ
measurements. It consists of a transmitter and a receiver creating a single path measurement
situation. The beam diameter is expanded to about 25 mm in order to improve performance in
high dust load (>1 g/m
Figure 5-2
1. Alignment screws
2. Purge tube
3. Purge flange
4. Wedge module, including wedge window
5. Lock rings (2 pc.)
6. Lens tube
7. Optics housing
8. Ground mount
9. Cable feed through - the transmitter has two (for hybrid cable and sensor connection cable)
and the receiver has one (for sensor connection cable)
10.Needle valve with fittings
The transmitter includes a lens and a connection for the optical fiber located in the focal plane
of the lens. The receiver contains a lens and the sensor electronics. The sensor electronics
consist of the detector module and the sensor card. The powering of the sensor electronics is
performed with a twisted pair copper cable delivering 24 VDC. The detector card is a detector
with pre-amplifier and the sensor card consists of driver electronics for the optical return signal
and DC/DC-converter.
The laser light is coupled from the optical fiber in the transmitter to its lens and passes through
a wedge tube (for protection of the lens) and then through the measurement volume. The lens
in the receiver (also protected with a wedge module) focuses the incoming laser light on the
detector where it is converted to an electrical signal. This signal is amplified and converted to an