High Pressure Flange; Lds 6, Optics Housing Cd 6 Ex - Siemens LDS 6 Operating Instructions Manual

In situ laser gas analyzers
Table of Contents

Advertisement

optical signal and returned to the central unit. The mounting of the optics housing requires a
flange with the dimensions DN65/PN6 or ANSI 4"/150 lbf.
5.4

High pressure flange

In some applications a high pressure window is needed to seal the process. These windows can
be purchased as spare parts and they exist in 3 versions namely, DN65/PN6, ANSI 4"/150lb and
a special version that will adapt the DN65/PN6 flange on the optics housing to a DN80/PN16
flange on the process - see the figure below.
Figure 5-3
5.5

LDS 6, Optics Housing CD 6 Ex

The Optics Housing CD 6 is also available in an Ex-version - see also separate user manual. This
has very low power electronics - intrinsically safe - and an IP65 enclosure.
LDS 6, Optics Housing CD 6
Operating Instructions, 01/2022, A5E01134377-05
High pressure window with DN80/PN16 towards the process and DN65/PN6 towards the
optics housing
Technical description
5.5 LDS 6, Optics Housing CD 6 Ex
31