Fluoroscopy Footswitch Configuration; Country-Specific Adaptation Of The Maximum Tube Current - Siemens ARCADIS Avantic SP Start-Up Instructions

Hide thumbs Also See for ARCADIS Avantic SP:
Table of Contents

Advertisement

Specific System Settings
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
Make country-specific changes and click <Save>.
Click <Home>.
If a change is made to the fluoro time, check this function.

Fluoroscopy footswitch configuration.

NOTE
Program the footswitch as follows.
Start service program and log in as described under "Service Login".
Select Main System / Configuration / Main System / Footswitch CFC / right.
Save the changes with <Save>.
Click <Home>.
If any reprogramming has been performed, the label attached to the footswitch must be
removed and replaced by the new label; the replacement label is attached to the
footswitch with a cable tie.
If a change is made to the pedal configuration, the function must be checked.

Country-specific adaptation of the maximum tube current

NOTE
Start service program and log in as described under "Service Login".
Siemens AG
Medical Solutions
Fluoroscopy alarm
4.5 min.
45
5 min.
50
The footswitch is set at the factory so that fluoroscopy is per-
formed with the left pedal and the freely-selectable operating
modes such as digital radiography, SUB, etc., are performed with
the right pedal. The setting can be changed to meet customer re-
quirements.
A maximum tube current limit is required - as far as known - only
in Denmark.
SPR2-330.815.01.01.02
02.05
Fluoroscopy blocking
5 min.
5
10 min.
50
Page 29 of 46
CS SD 24
29