Specific System Settings
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
•
Make country-specific changes and click <Save>.
•
Click <Home>.
•
If a change is made to the fluoro time, check this function.
Fluoroscopy footswitch configuration.
NOTE
Program the footswitch as follows.
•
Start service program and log in as described under "Service Login".
Select Main System / Configuration / Main System / Footswitch CFC / right.
•
Save the changes with <Save>.
Click <Home>.
If any reprogramming has been performed, the label attached to the footswitch must be
removed and replaced by the new label; the replacement label is attached to the
footswitch with a cable tie.
If a change is made to the pedal configuration, the function must be checked.
Country-specific adaptation of the maximum tube current
NOTE
•
Start service program and log in as described under "Service Login".
Siemens AG
Medical Solutions
Fluoroscopy alarm
4.5 min.
45
5 min.
50
The footswitch is set at the factory so that fluoroscopy is per-
formed with the left pedal and the freely-selectable operating
modes such as digital radiography, SUB, etc., are performed with
the right pedal. The setting can be changed to meet customer re-
quirements.
A maximum tube current limit is required - as far as known - only
in Denmark.
SPR2-330.815.01.01.02
02.05
Fluoroscopy blocking
5 min.
5
10 min.
50
Page 29 of 46
CS SD 24
29
0
0
ARCADIS Avantic