Testing The Protective Conductors - Siemens ARCADIS Avantic SP Start-Up Instructions

Hide thumbs Also See for ARCADIS Avantic SP:
Table of Contents

Advertisement

General Startup

Testing the Protective Conductors

The system must be free of voltage.
Measuring de-
vice:
Measuring de-
vice:
Test procedure:
Siemens AG
Medical Solutions
Protective conductor testing device; e.g., safety tester Unimed
1100
Max. 0.2 Ohm (observe country-specific regulations!) (current:
10 A)
Measure between all touchable, conductive parts of the system
and the protective conductor wire at the ARCADIS Avantic
power plug.
SPR2-330.815.01.01.02
02.05
Page 19 of 46
CS SD 24
19
0