Oscilloscope And Probes; Software; Measurement - Nvidia Jetson AGX Orin Series Compliance Manual

Table of Contents

Advertisement

Oscilloscope and Probes

The oscilloscope and its probes are used to measure and display the signals required for the
RGMII compliance testing. Since the interface runs with a maximum frequency of up to 125
MHz, any oscilloscope, and probes with a bandwidth of more than 625 MHz are acceptable.
Note: Scope, probes, and test equipment should be calibrated and de-skewed, and must meet
industry standards to ensure accurate measurements.

Software

For creating traffic on the Orin RGMII interface, software (for example, iperf, Ping, SCP) or
traffic generator hardware is needed.
To access the registers, depending on the software platform, devmem2 or in32/out32 may be
used.
Linux environment example:
# devmem2 0x02318804
# devmem2 0x02318804 w 0xA0010000
QNX environment example:
# in32
0x02318804
# out32 0x02318804 0xA0010000

Measurement

Measurement consists of the following steps:
Attach the oscilloscope probes to the DUT, as close as possible to the Ethernet PHY (or
1.
Switch) receiver interface.
Power on the DUT.
2.
Configure the DUT to drive traffic.
3.
Observe the electrical measurements on the oscilloscope.
4.
Note: Ideally all Orin module RGMII transmitter signals should be examined. However, to save
time and effort, only the clock (module RGMII_TXC, SoC GP147_RGMII0_TXC) and one data line
(for example, module RGMII_TD0 – SoC GP148_RGMII0_TD0) may be monitored.
PRELIMINARY INFORMATION
Jetson AGX Orin Series Tuning and Compliance Guide
Ethernet Compliance Test Guide
// read the register 0x02318804
// write 0xA0010000 as a word to
0x02318804
// read the register 0x02318804
// write 0xA0010000 as a word to
0x02318804
DA-11040-001_v0.7 | 44