Upgrading an HP 1100 Series LC
Thermo Scientific
The Xcalibur 2.0 and higher data system supports control of the Hewlett-Packard (HP)
1100 Series LC only if you upgrade the HP 1100 communication interface to an Ethernet
interface.
To upgrade your HP 1100 Series LC
1. Order the appropriate upgrade parts from your local office for Thermo Electron San Jose
products:
• If the HP 1100 is not interfaced to an MS detector as an Xcalibur-controlled GPIB
inlet device, order the JetDirect Ethernet Control kit, which contains the parts listed
in
Table
4.
• If the HP 1100 is interfaced to an MS detector as an Xcalibur-controlled GPIB inlet
device, order the parts listed in
Note An HP 1100 LC system that is currently interfaced to a MS detector as an
Xcalibur-controlled GPIB inlet device should have already installed the contact
closure PCB and the external contact closure cable. Therefore, you need to order
only the parts specified in
Table 4. JetDirect Ethernet Control Kit (P/N OPTON 30018)
Part Number
00012-27714
00012-27716
00012-70008
00825-01-00024
00825-01140
Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition
Table
5.
Table
5.
Description of Kit
Contact closure PCB
External contact closure cable
Ethernet 10 Base-T cable (2)
10/100 Autosensing 8-port Ethernet switch
HP JetDirect 400N PCB