6 - 8
Place the Videomed DC in the zero position on the Compact optics (6/Fig. 2) and clamp it
in place.
Connect the cable from the Videomed DC to the Compact optics and secure it with cable
ties.
Make sure that the cable does not protrude past the edge of the board and that it lies in
the recess of the Videomed DC board.
Install the new control section included in the I.I. delivery volume.
Connect cables E1, E2.
Secure the new I.I. to the mounting with the screw.
- The I.I. can be swivelled slightly to the side to permit easier cable routing.
Route the cables for the high voltage section back through the recess in the I.I.
(3/Fig. 1 or 3 Fig. 4).
Swivel the I.I. back again.
- When doing this, make sure that all the cables are correctly positioned.
Secure the I.I. on the mounting (5/Fig. 2).
The C-arm can now be rotated to permit easier installation of components (I.I. above,
tube assembly underneath).
Reinstall the O-ring and the cap nut on the anode cable.
- The cable length from the beginning of the white wire to the cap nut should be 70 mm.
Reinstall the high voltage section in the C-arm.
- Reconnect all the cables.
Reinstall the hand rail above the Videomed DC.
Reinstall both jacks D4.X1 for the Videomed DC and X45 for the cassette contact on the
receptacle.
Reconnect the D4.X1 plug for the Videomed DC and the X45 plug for the cassette
contact.
Reconnect plug M30.X1.
Reconnect plug D120.X1 to the control section of the I.I. mini power supply.
Reconnect the protective conductor to the ground screw (5/Fig. 1).
Reinstall the cable clips (4/Fig. 1).
Reinstall the counterbalance weight for the 7"/17cm I.I
Check the existence of the sealing ring 6 /Fig. 3.
NOTE
SIREMOBIL Compact
System Manual
Replacing boards / replacing components
Check the presence of the sealing ring 6/Fig. 3 at the new I.I.
If none is present, remove the sealing ring from the old I.I. (glued
to the I.I. housing with 4 to 6 dots), lay it in the groove of the new
I.I. housing and fix it with 4 to 6 dots of adhesive
(e.g. superglue).
Register 5
RXR2-130.061.02
Rev. 03
Page 8 of 18
10.03
CS PS 24
Siemens AG
Medical Engineering