Adjustments / Programs
Checking the programmed values
NOTE
Prerequisites
The Qt * Kg value was checked and is programmed. Refer to Programming.
•
Connect the service PC to the serial interface on board D1.
•
Start the service program.
•
Select the Adjustment, Calibrations menu.
•
Click "Yes" to answer the query.
•
In the Combo box "Function Group:", select "Diamentor".
•
In the window "Available Functions", select the line "Start Diamentor Test" and click on
"Execute" .
•
The tested Qt*Kg value is displayed on the monitor of the SIREMOBIL Compact.
•
This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: +/- 5%.
•
By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
•
The potentiometer CAL on the electronic system of the area dose product measurement
device can be used to adjust the test value to the calculated and programmed Qt*Kg
value.
•
After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
•
The test is ended.
•
Check the accuracy of the area dose product measurement device. Refer to the next
page.
Siemens AG
Medical Engineering
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.
Register 5
RXR2-130.061.02
Rev. 03
Page 15 of 22
10.03
CS PS 24
5 - 15
5
SIREMOBIL Compact
System Manual