Siemens SIREMOBIL Compact Installation And Setting Instructions Manual
Siemens SIREMOBIL Compact Installation And Setting Instructions Manual

Siemens SIREMOBIL Compact Installation And Setting Instructions Manual

Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

Quick Links

Attaching the area dose measurement device
Register 3
Print No.: RXR2-130.033.01.04.02
Replaces: RXR2-130.033.01.03.02
© Siemens AG 1996
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 04.02

Advertisement

loading

Summary of Contents for Siemens SIREMOBIL Compact

  • Page 1 Attaching the area dose measurement device © Siemens AG 1996 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3 96 60 051 RE999 Tools required • Tool kit (refer to ARTD / part 3) • 1 set of Allen keys Siemens AG Register 3 RXR2-130.033.01 Page 3 of 10 SIREMOBIL Compact Medical Engineering Rev. 04 04.02 CS PS 24...
  • Page 4: Installation

    Install measurement board D16 according to Fig. 2. • Plug the two connectors (2/Fig. 2) into PC board D16. • Connect the enclosed ribbon cable to D1.X9 and D16 (3/Fig. 2). SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 4 of 10 Siemens AG Rev.
  • Page 5 Installation and Setting Instructions Memoskop Programming SIREMOBIL Compact starting with serial number 2000 After selecting the "Disable dose rate selection in User Set Up" NOTE parameter, the "Storage of dose on disk" and "Dose on hardcopy output" parameters become available in the Memoskop User Setup and visible to the customer.
  • Page 6 Programming and adjusting the area dose measurement device This adjustment is used to calibrate the area dose measurement NOTE device of the SIREMOBIL Compact display. Calculating the CONSTANT parameter (Qt * Kg) • Read and record the Kg value shown on the electronics unit of the area dose measurement device.
  • Page 7 Press 2 times the key ->0<- at the control panel of the basic unit (reset key for fluoroscopy timer). • The tested Qt*Kg value is displayed at the monitor of the SIREMOBIL Compact. • This value must match the programmed CONSTANT (Qt*Kg) value.
  • Page 8 Calculating the area dose product • To calculate the measured area dose product: (measured area dose product FDPg) = (measured dose K ) * (average diameter) 0.785. SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 8 of 10 Siemens AG Rev. 04 04.02...
  • Page 9 For deviations > 0.3, replace the area dose measurement device. Sirephos Sirephos Meas. chamber Centering cross I.I. I.I. Fig. 3 Fig. 4 Fig. 5 Siemens AG Register 3 RXR2-130.033.01 Page 9 of 10 SIREMOBIL Compact Medical Engineering Rev. 04 04.02 CS PS 24...
  • Page 10: Final Steps

    Measuring devices actualized. Page 6 Corrections in descriptions added. Page 7 Corrections in descriptions added. Page 9 Maximum values for the protective conductor resistance added. SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 10 of 10 Siemens AG Rev. 04 04.02 CS PS 24...