WARNING
CAUTION
CAUTION

To Install the Switch Module

To prevent electrical shock, turn off the mainframe and remove the power
cable before starting the instruction.
Be careful about the module pins used for internal connection to the Agilent B2200.
The pins can be damaged easily.
Use clean handling and anti-static procedures when removing, configuring, and
installing the switch modules. The modules contain components that can be
damaged by static electricity.
The following procedure explains the module installation and removal.
1. Turn off the Agilent B2200, then wait at least 10 seconds before you remove or
install a module.
2. Loosen the screws on both the left and right edges of a blank panel or a module
attached to the slot you want to install a new module.
3. Remove the blank panel or the module.
4. Align the new module with the left and right slot guide rails. Then the
component side should be facing up.
5. Push the module into the slot until you feel it seat firmly into the connector at the
back of the slot.
6. Screw in the screws on the left and right edges of the module.
7. Execute the self-test and the relay test. See "Self-Test" on page 2-10.

To Install the Blank Panel

To prevent thermal damage to the module, be sure that blank panels are installed in
all unused slots.
If the blank panel is not installed to cover an unused slot, install the blank panel as
follows:
1. Align the blank panel over the unused slot.
2. Screw in the screws on the left and right edges of the blank panel.
Agilent B2200 User's Guide, Edition 2
Installation
Installing the B2200
2-9