Set Up a New Radar Level Sensor

Level sensors need to have three settings setup after
installing on the tank of the PFP machine. The tank
must be empty, or the level needs to be at least below
the bottom of the installed probe with the tank cover
closed.
1. Open the sensor head cover over the display:
a. Loosen the securing clamp with an Allen
wrench (if clamp is present).
b. Unscrew the round cover and pull away.
3A8047B
NOTE: Use the operating keys to navigate within the
operating menu and to select options from a list.
2. Turn power on the machine. Set EMPTY CAL, then
3. Set EMPTY CAL to .666 meters:
ti24380a
Key
Meaning
"Minus" Key
• In a selection list, this key will move
the selection bar upward.
• In an input matrix, this key will
move the selection bar backward.
"Plus" Key
• In a selection list, this key will move
the selection bar downward.
• In an input matrix, this key will
move the selection bar forward.
"Enter" Key
• Opens the marked sub-menu or
E
parameter.
• Confirms a changed parameter
value.
"Escape" Key combination (press
keys simultaneously)
+
• Closes a parameter without
accepting the changes.
• Quits the current menu layer and
returns to the next higher layer.
FULL CAL, then do a mapping of the Empty Tank.
a. Press E, then +, +, to get to Set Up. Then press
E to get to EMPTY CAL, then press E.
b. Press + until decimal point (.) appears, then
press E.
c. Press + until 6 appears, then press E.
d. Press + until 6 appears, then press E.
e. Press + until 6 appears, then press E.
f.
Press and hold E until the hourglass appears,
then release.
Repair
27