About this document Prosonic M FMU44 HART About this document Document function These Operating Instructions provide all of the information that is required in the various phases of the life cycle of the device. Symbols 1.2.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter serial number from nameplate • Endress+Hauser Operations app: Enter serial number from nameplate. 1.3.1 Standard documentation Technical Information (TI) Planning aid ‒...
The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. Endress+Hauser...
The device meets the legal requirements of the applicable EAC Directives. These are listed in the corresponding EAC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the EAC mark. IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions.
‣ Enter the serial number from the nameplate in the Endress+Hauser Operations app or scan the 2-D matrix code on the nameplate with the camera All of the information on the measuring device and on the scope of the technical documentation pertaining to the device is displayed.
Prosonic M FMU44 HART Mounting Place of manufacture: See nameplate. Storage and transport • Pack the measuring device so that it is protected against impact. The original packaging offers the best protection • Permitted storage temperature: –40 to +80 °C (–40 to 176 °F) 4.4.1...
Page 10
Mounting Prosonic M FMU44 HART • Only mount one device per vessel: signals from multiple devices mutually affect one another • Determine the detection range using the 3-dB beam angle α 5.1.2 Beam angle W = 2 · D · tan A0038221 ...
Page 11
Prosonic M FMU44 HART Mounting 5.1.4 Installation conditions for the sensor for flow measurements • Mount the measuring device on the upstream side immediately above the maximum upper water level H • Take the blocking distance into account • Position the measuring device in the center of the flume or weir •...
Page 12
Mounting Prosonic M FMU44 HART ≥3 H ≥ 2 H A0038223 8 Triangular weir (example) BD Blocking distance of the sensor Empty calibration (enter during commissioning) Full calibration Maximum upstream level 5.1.5 Mounting examples A0038822 9 Mounting examples Mounting with universal flange (e.g.
Prosonic M FMU44 HART Mounting 5.1.6 Turning the housing 4 mm A0037777 10 Turning the F12 or T12 housing, (F12 housing in example) 1. Loosen the securing screw. 2. Rotate the housing in the desired direction, maximum 350 °.
Page 14
Mounting Prosonic M FMU44 HART A0038238 11 Parameter for the correct operation of the device BD Blocking distance Safety distance Empty calibration Measuring span Mounting nozzle A0038371 12 Definition of nozzle length L and nozzle diameter D Nozzle diameter...
Prosonic M FMU44 HART Electrical connection In the In safety dist. (016) function define how the device should react if the safety distance is undershot. Options and meaning • Warning The measuring device displays an error message but keeps measuring.
Electrical connection Prosonic M FMU44 HART Connecting the measuring device 6.2.1 Connection in the F12 housing 1 → 2 → 3 → A0038240 A0038241 A0038242 ‣ ‣ ‣ Unscrew the cover Remove the display if one is Remove the cable from the...
Prosonic M FMU44 HART Electrical connection 6.2.2 Connection in the T12 housing 1 → 2 → 3 → A0038256 A0038257 A0038258 ‣ ‣ Unscrew the cover Avoid moisture in the housing, provide a loop to allow moisture to drain off 4 →...
Electrical connection Prosonic M FMU44 HART ‣ A standard device cable suffices if only the analog signal should be used. Use a shielded cable if working with a superimposed communication signal (HART). Protective circuits against reverse polarity, RF influences and overvoltage peaks are integrated in the device.
Prosonic M FMU44 HART Operation options A0038292 14 Position of outer ground terminal, example of F12 housing • Connect the potential equalization line to the outer ground terminal of the transmitter • In the case of hazardous area applications, only ground on the sensor side •...
Operation options Prosonic M FMU44 HART The first two digits identify the function group, examples: • Basic setup: 00 • Safety settings: 01 • Linearization: 04 The third digit refers to the individual functions within the function group, examples: • Tank shape: 002 •...
Page 21
Prosonic M FMU44 HART Operation options A0047446 18 Envelope curve display Envelope curve, view Symbols ALARM_SYMBOL The device is in an alarm state. If the symbol flashes, this indicates a warning. LOCK_SYMBOL Device is locked. Additional entries are not possible.
Operation options Prosonic M FMU44 HART 7.3.2 Operating the local display Group selection 1. Press Change from the Measured value display to the Group selection. 2. Press or : select the Function group. 3. Press : confirm your choice.
Operating Instructions BA00060S/04. 8.1.2 Operation with FieldCare FieldCare is an Endress+Hauser asset management tool based on FDT technology. FieldCare configures all Endress+Hauser devices and devices from other manufacturers that support the FDT standard. Hardware and software requirements: www.de.endress.com - Search: FieldCare FieldCare supports the following functions: •...
Page 24
System integration Prosonic M FMU44 HART Connection options: • HART via Commubox FXA195 and the USB port on a computer • Commubox FXA291 with ToF Adapter FXA291 via service interface Menu-guided commissioning 20 View in the FieldCare operating program •...
Prosonic M FMU44 HART Commissioning Commissioning The Commissioning section contains the following steps: • Function check • Switching on the measuring device • Establishing a connection via Fieldcare • Configuring the measuring device • Displaying the envelope curve Function check 9.1.1...
Commissioning Prosonic M FMU44 HART Parameters that need to be configured when the device is switched on for the first time: • Language Select the language of the display. • Length unit Select the unit of length in which the distance is measured.
Page 27
Prosonic M FMU44 HART Commissioning Configuration of the measuring point Tank shape function (002) Selection A0038388 23 Options in the "Tank shape" function Dome ceiling Horizontal cyl Bypass or stilling well/ultrasound guide pipe No ceiling, e.g. dumps, open levels, basins, flume...
Page 28
Commissioning Prosonic M FMU44 HART A0038402 24 Some of the options in the "Process cond." function Calm surface Turb. surface Add. agitator Fast change Standard solid Solid dusty Conveyor belt Not in graphic: Standard liq. and Test: no filter Description of options •...
Page 29
Prosonic M FMU44 HART Commissioning • Solid dusty • For dusty bulk solids • The filters are set in such a way that even relatively weak wanted signals are detected • Conveyor belt • For bulk solids with rapid changes in the level, e.g. conveyor belts •...
Page 30
Commissioning Prosonic M FMU44 HART NOTICE If the blocking distance is undershot, this may cause a device malfunction. ‣ Mount the measuring device high enough that the blocking distance is not reached at the maximum fill level. ‣ Define the safety distance (SD).
Prosonic M FMU44 HART Commissioning Range of mapping (052) function • The range of mapping is displayed, confirm the proposed value or enter your own value • The sensor membrane is always the reference point • The value can be edited by the user •...
Page 32
Commissioning Prosonic M FMU44 HART 9.5.3 Envelope curve display (0E3) function If cyclic envelope curve display is active on the display, the measured value is updated at a slower cycle time. Recommendation: exit the envelope curve display after optimizing the measuring point. Press (the device does not leave the envelope curve display automatically.)
Prosonic M FMU44 HART Diagnostics and troubleshooting Activating the horizontal zoom mode 1. Press or . is displayed. 2. increases the horizontal scale. 3. reduces the horizontal scale. Activating the move mode 1. Then press .
Diagnostics and troubleshooting Prosonic M FMU44 HART 10.3 Types of error Alarm (A), symbol is lit continuously The output signal adopts a value that can be specified by the Output on alarm (010) function. • MAX: 110 %, 22 mA • MIN: -10 %, 3.8 mA •...
Prosonic M FMU44 HART Diagnostics and troubleshooting W103 = Initializing If the message does not disappear, replace the electronics W153 = Initializing • Wait a few seconds • If the error remains active, switch the voltage off/on W601 = Linearization curve not monotone...
Page 36
Diagnostics and troubleshooting Prosonic M FMU44 HART 100% t → A0038459 Expected result Error occurred 1. In the Tank shape (002) function, select the corresponding option for measurements in a bypass or ultrasound guide pipe. 2. Perform mapping. Error: No change of measured value when filling or emptying.
Page 37
Prosonic M FMU44 HART Diagnostics and troubleshooting 100% t → A0038462 Expected result Error occurred 1. Perform interference echo suppression (mapping). 2. In the Process cond. (004) function, select "Turb. surface" or "Add. agitator". 3. Increase the output damping (058).
Diagnostics and troubleshooting Prosonic M FMU44 HART 100% E641 t → A0038464 Expected result Error occurred 1. Check application parameters (002), (003) and (004). 2. Select another installation position. 3. Select a larger sensor. 4. Align the sensor so that it is parallel to the surface of the product, particularly for bulk solids applications.
Repair 12.1 General notes Endress+Hauser repair concept • Measuring devices have a modular design • Customers can carry out repairs For more information on service and spare parts, please contact your Endress+Hauser sales representative. Endress+Hauser...
WARNING Incorrect repair can compromise electrical safety! Explosion Hazard! ‣ Only specialist personnel or the Endress+Hauser service team may carry out repairs on Ex-certified devices. ‣ Relevant standards and national regulations on hazardous areas, safety instructions and certificates must be observed.
The parameters are recorded and transmitted using Fieldgate FXA42, FXA30 and FXA30B gateway types. SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the Endress+Hauser portal, the user is provided with the data over the Internet.
Prosonic M FMU44 HART Technical data Order number: 71214483 For details, see "Technical Information" TI01229S and Operating Instructions BA00050S. 13.2.5 SupplyCare Enterprise SCE30B Inventory management software that visualizes levels, volumes, masses, temperatures, pressures, densities or other tank parameters. The parameters are recorded and transmitted using the Fieldgate FXA42 gateway type.
Technical data Prosonic M FMU44 HART 20 mA 100% 4 mA A0038386 31 Parameters for the correct operation of the device BD Blocking distance Safety distance Empty distance Level Distance from the sensor membrane to the surface of the product Measuring span (full distance) 14.1.2...
Prosonic M FMU44 HART Technical data Standard • Current consumption 4 mA | terminal voltage 14 to 36 V • Current consumption 20 mA | terminal voltage 8 to 36 V Ex ia • Current consumption 4 mA | terminal voltage 14 to 30 V •...
Technical data Prosonic M FMU44 HART 14.4.2 Response time The response time depends on the configured application parameters. Minimum values: Two-wire version: ≥ 3 s 14.4.3 Measured value resolution 2 mm (0.08 in) 14.4.4 Measuring error Better than ±4 mm (±0.16 in)or 0.2 % of the measuring distance, the larger value applies Observe typical measured error under reference operating conditions •...
Prosonic M FMU44 HART Appendix 14.5.2 Storage temperature –40 to +80 °C (–40 to +176 °F) 14.5.3 Climate class • DIN EN 60068-2-38 (Test Z/AD) • DIN/IEC 68 T2-30Db 14.5.4 Degree of protection When housing is closed, tested according to •...
Page 48
Appendix Prosonic M FMU44 HART • Empty calibration 005 • Full calibration 006 • Distance/measured value 008 Safety settings 01 • Outp. on alarm 010 • Outp. on alarm 011 • Outp. echo loss 012 • Ramp %span/min 013 • Delay time 014 •...
Page 49
Prosonic M FMU44 HART Appendix Display 09 • Language 092 • Back to home 093 • Format display 094 • No. of decimals 095 • Sep. character 096 • Display test 097 Diagnostic 0A • Present error 0A0 • Previous error 0A1 •...