4
1-

General information

1 General information and safety guidelines

Validity of the document

This document is valid for the subsequent installation of a multi-function footswitch in an
ARCADIS Varic system or an ARCADIS Orbic system that has been prepared for the
multi-function footswitch option.

Requirements

The ARCADIS Varic and ARCADIS Orbic systems delivered after December, 2006
(approximately) have already been prepared for the installation of the multi-function foot-
switch. The following are indications that a system has been prepared for this installation:
NOTE

Required material

Multi-function footswitch option, material number 08079548

Tools and auxiliary equipment required

Standard tool kit
Cable ties (included in the delivery volume of the multi-function footswitch)
ARCADIS Varic/Orbic
General information and safety guidelines
There is an externally accessible footswitch plug connector socket on the basic
unit.
The holes for attaching the footswitch holder to the cover panel have already
been drilled and are fitted with dummy covers.
The internal wiring for the footswitch plug connection has been prepared for the
upgrade to the multi-function footswitch.
Software version VB13C or higher is installed on the system.
It is not possible to upgrade ARCADIS Varic or ARCADIS Orbic
systems with the multi-function footswitch option upgrade kit
unless these preparations have been made.
SPR2-310.812.03.01.02
09.06
Page 4 of 20
CS PS SP
0
0
0
0
0
Siemens AG