Final work steps

Covers

Close all cover panels if any are still open. Make sure the protective conductors make
good contact.

Protective conductor resistance test

Observe the instructions for the protective conductor resistance test in Chapter 1 of
these instructions.
Perform the protective conductor resistance test according to ARTD-002.731.17...

Customer training

NOTE
Explain the functions of the multi-function footswitch to the customer. Also, demon-
strate how to place the footswitch in the holder.
ARCADIS Varic/Orbic
Function check and final work steps
The protective conductor resistance must not exceed 0.2 ohms.
The system operating manual already describes the multi-func-
tion footswitch and its functions.