20
Final work steps
Covers
•
Close all cover panels if any are still open. Make sure the protective conductors make
good contact.
Protective conductor resistance test
•
Observe the instructions for the protective conductor resistance test in Chapter 1 of
these instructions.
•
Perform the protective conductor resistance test according to ARTD-002.731.17...
Customer training
NOTE
•
Explain the functions of the multi-function footswitch to the customer. Also, demon-
strate how to place the footswitch in the holder.
ARCADIS Varic/Orbic
Function check and final work steps
The protective conductor resistance must not exceed 0.2 ohms.
The system operating manual already describes the multi-func-
tion footswitch and its functions.
SPR2-310.812.03.01.02
09.06
Page 20 of 20
CS PS SP
3.1
0
0
0
Siemens AG
Medical Solutions