Function check and final work steps
3-
Function check
3Function check and final work steps
•
Plug the monitor cable into the basic unit.
•
Switch on the system and wait for it to boot.
•
Register an emergency patient.
•
X
Briefly release fluoro with the fluoro pedal.
•
On the multi-function footswitch, press the "Save Image" button.
•
On the multi-function footswitch, press the "Select Mode" button several times.
NOTE
•
With the "Select Mode" button, select the PFC mode.
•
X
In the selected mode, briefly press the "Release Radiation" foot pedal.
•
Select each mode in turn and check the function of each.
•
X
Once again, briefly release fluoro with the fluoro pedal.
Siemens AG
Medical Solutions
Fluoro mode is executed and radiation is released.
The LIH image is saved, and is also displayed on the right-hand monitor.
The PFC, subtraction, roadmap, and DR modes are selected successively, and
the LED for the currently selected mode lights up. If the button is pressed again
after DR has been selected, the modes are successively selected again in the
order: PFC, subtraction, roadmap, DR.
The subtraction and roadmap modes can only be selected if the
license for subtraction has been installed.
The selected mode is executed and radiation is released.
Fluoro mode is executed and radiation is released.
During the release of radiation, the fluoro button LED on the operating console of
the basic unit lights up.
After the radiation release stops, the previously selected mode is selected again
automatically, and the LED on the corresponding button (PFC, SUB, ROAD or
DR) lights up.
SPR2-310.812.03.01.02
09.06
Page 19 of 20
CS PS SP
19
0