Mechanical installation and wiring
•
Wire the modified Y cable as follows:
•
Attach the ends of the Y cable along the existing cable assemblies using the cable ties
included in the accessory kit.
•
Insert the plug for the multi-function footswitch into the footswitch socket and lock it in
place.
Closing the covers
•
Reattach the ground wire to the rear cover panel of the basic unit. Make sure the wire
has good contact with the cover panel.
•
Reattach the rear cover panel to the basic unit.
Siemens AG
Medical Solutions
Plug the Y cable's D31.X2 plug into the new D31.X2 board.
Plug the Y cable's D31.X5 plug into the new D31.X5 board.
Plug the Y cable's D1.X14 plug into the D1.X14 board.
Plug the Y cable's X21 plug into the open X21 socket on the footswitch mounting
jack.
SPR2-310.812.03.01.02
09.06
Page 15 of 20
CS PS SP
15
0
ARCADIS Varic/Orbic