Mechanical installation and wiring
Fig. 8:
Insert the threaded ends of the three spacer bolts from above into the holes in the frame
of the basic unit. The spacer bolts are visible in
ing plate..
Screw down the three spacer bolts from the underside of the basic unit's frame with
three nuts, washers, and lock washers.
Attach the 4 circuit board holders to the mounting plate with 4 screws, 4 washers, and
retaining rings. See
Clip the D31 board onto the circuit board holders. Make sure the board clicks in all the
way in the circuit board holders.

Wiring for systems without the 3D reconstruction option installed

Removing the wire from the M16.K1.3 and M16.K1.4 relays to D1.X14
Fig. 9:
At the M16.K1 relay (emergency stop relay, see
and M16.K1.4 wires.
On the D1 board, detach the D1.X14 plug.
Remove the wire. To do so, cut open the appropriate cable ties.
Siemens AG
Medical Solutions
D31 holder and board, installation location
(Fig. 7 / p.
M16.K1 relay location
The wire will be replaced with the Y cable included in the accessory kit.
SPR2-310.812.03.01.02
09.06
(Fig. 8 / p. 13)
12).
(Fig. 9 / p.
Page 13 of 20
CS PS SP
underneath the mount-
13)) remove the M16.K1.3
ARCADIS Varic/Orbic
13