Wiring For Systems With The 3D Reconstruction Option Installed - Siemens ARCADIS Varic Installation Instructions Manual

Multi-function switch
Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

14
Wiring the Y cable
Fig. 10: X21 plug location
Wire the Y cable from the accessory kit as follows:
Attach the ends of the Y cable along the existing cable assemblies using the cable ties
included in the accessory kit.
Insert the plug for the multi-function footswitch into the footswitch socket and lock it in
place.

Wiring for systems with the 3D reconstruction option installed

Wiring the Y cable
Modify the Y cable from the accessory kit as follows:
NOTE
ARCADIS Varic/Orbic
Mechanical installation and wiring
Plug the Y cable's D31.X2 plug into the new D31.X2 board.
Plug the Y cable's D31.X5 plug into the new D31.X5 board.
Plug the Y cable's RNH.3 plug into the M16.K1.3 relay.
Plug the Y cable's RNH.4 plug into the M16.K1.4 relay.
Plug the Y cable's D1.X14 plug into the D1.X14 board.
Plug the Y cable's X21 plug into the open X21 socket, connected to the foot-
switch mounting jack. See
Locate the wires labeled RNH.3 and RNH.4. These lead to the Y cable's D1.X14
plug.
Cut the RNH.3 and RNH.4 wires at a distance of about 8cm from the D1.X14 plug,
and insulate them with heat-shrink tubing. The label with the designation D1 X14
should still be in place and should remain so.
The RNH.3 and RNH.4 wires are not needed when the 3D recon-
struction option is installed.
SPR2-310.812.03.01.02
09.06
(Fig. 10 / p.
14).
Page 14 of 20
CS PS SP
0
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Arcadis orbic