Detector Positionning (For Version With Removable Detector) - AGFA D2RS User Manual

Digital remote controlled table with dynamic flat panel detector
Table of Contents

Advertisement

2.2. DETECTOR POSITIONING (for version with removable detector)
The version of remote controlled table with removable detector has a slightly different interface, because the main
screen display icons that concern this detector.
CAUTION: Before any table movement, chek that the detector is in right place for the exam, physically
inserted in the table or free in room, and by software that will indicate the following on the first screen
Digital Remote Controlled Table with dynamic flat panel detector
The detector is in place, inserted, in landscape format.
HEAD table side
You can change the orientation by pressing the icon
The detector is in place, inserted, in portrait format.
You can change the orientation by pressing the icon
The detector is inserted, but its orientation is not detected.
Please, check if there is any detector positioning trouble firstly by pressing again the icon,
and if nothing happens, remove the detector, insert it again. If the same icon appears,
press again the icon. If nothing happens, call After-Sales Service
The detector is not detected in place, inserted. Please, correct this situation verifying
that the detector is there and correctly inserted.
User's Manual
FOOT table side