Measuring Temperature Using Resistance Temperature Detectors (Rtds) - Fluke 525B User Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525B:
Table of Contents

Advertisement

Measuring Temperature Using Resistance Temperature Detectors (RTDs)

You must use a 4-wire connection to achieve the Calibrator accuracy
specifications.
To measure the RTD output from an instrument:
1. Connect the Calibrator to the instrument you want to measure as shown in
Figure 3-3.
2. Press  to select RTD mode.
3. If necessary, press   to put the Calibrator in RTD IN mode.
4. Press  to select the desired RTD type.
You can toggle between ºC and ºF temperature units by pressing
 .
725
MULTIFUNCTION CALIBRATOR
V mA
ZERO
LOOP
V mA
Hz
MEAS
SOURCE
TC
RTD
°
C
°
F
STORE
100%
SETUP
25%
RECALL
25%
0%
Figure 3-3. Measuring RTD Output from an Instrument
To measure temperatures using an RTD probe:
1. Connect an RTD probe to the Calibrator as shown in Figure 3-4.
2. Press  to select the RTD mode.
3. If necessary, press   to put the Calibrator in RTD IN mode.
VOLTS
mA
RTD
100mA MAX
HI
HI
OUTPUT
100V MAX
TC
LO
LO
INPUT/OUTPUT
20V PK
MAX
4W RTD
HI
20V PK
MAX
INPUT
SENSE
LO
20V PK
MAX
Using Input Mode
Using Input Mode
STBY
VOLTS
TC
TYPE
OPR
mA
RTD
UNITS
OUTPUT
INPUT
ZERO
7
8
9
SHIFT
SETUP
CJC
C / F
4
5
6
SET
RECALL
AUTOSET
2
3
1
RNG LOCK
LOCAL
EXP
/
0
ENTER
3
CE
fcn06f.eps