Fluke 525B User Manual page 80

Temperature/pressure calibrator
Hide thumbs Also See for 525B:
Table of Contents

Advertisement

525B
Users Manual
OUT
(Output command) Sets the output of the Calibrator and establishes a new
reference point for the error mode. To source or measure a temperature, select the
desired sensor and sensor parameters first. (See the TSENS_TYPE, RTD_TYPE,
and TC_TYPE commands.)
Use multipliers, k, m, u with the OUT command, as desired.
Parameters: <value> V
Examples:
Each example shows a value and unit, e.g., −15.2 V. If a value is entered without a
unit, the value of the existing output is changed, when logically allowed.
OUT?
(Output query) Returns the output amplitude of the Calibrator. Multipliers (e.g., K
or M) are not used in the response.
Response:
Examples:
The respective values for the above examples are:
5-10
x
x
IEEE-488
RS-232
<value> A
<value> OHM
<value> CEL
<value> FAR
OUT 15.2 V
OUT 1.2 mA
OUT 5 Ohm
OUT 100 CEL
OUT −32 FAR
x
x
IEEE-488
RS-232
<amplitude value>,<units>
OUT? returns −1.520000E+01,V
OUT? returns 1.88300E−02,A
OUT? returns 1.23000E+00,V
OUT? returns 4.00000E+03,OHM
OUT? returns 1.52000E+01,V
OUT? returns 1.0430E+02,CEL
−15.2 V
18.83 mA
1.23 V
4 KΩ
15.2 V
104.3 °C
x
x
Sequential
Overlapped
Volts dc
Current dc
Resistance
Temperature (Celsius)
Temperature (Fahrenheit)
(volts; 15.2 V)
(current; 1.2 mA)
(ohms; 5 Ω)
(temperature; 100 °C)
(temperature; −32°F)
x
x
Sequential
Overlapped