Fluke 525B User Manual page 90

Temperature/pressure calibrator
Hide thumbs Also See for 525B:
Table of Contents

Advertisement

525B
Users Manual
*WAI
(Wait-to-Continue command) Prevents further remote commands from being
executed until all previous remote commands have been executed. For example, if
you send an OUT command, you can cause the Calibrator to wait until the output
has settled before continuing on to the next command if you follow OUT with a
*WAI command. The *WAI command is useful with any overlapped command,
preventing the Calibrator from processing other commands until the overlapped
command is processed.
Example:
Process all existing commands before continuing.
ZERO_MEAS
(Zero Measure command) Zeros the pressure module and the TC mV offset. See
Chapter 3 for the TC mV zero calibration procedure.
Parameters: <value> in current pressure units for absolute pressure modules.
Example:
ZERO_MEAS?
(Zero Measure query) Returns the zero offset for absolute pressure modules.
Response:
Example:
5-20
x
x
IEEE-488
RS-232
*WAI
x
IEEE-488
ZERO_MEAS
x
IEEE-488
<zero offset>,<units>
ZERO_MEAS? returns −1.520000E+00,PSI
x
x
Sequential
Overlapped
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
Overlapped
x
Overlapped