Table 1-4. Function Keys (cont)
Press  to advance through the LCD Backlight, Interface, and
Address menus.
• Use  and  to adjust LCD backlight when the LCD menu is
displayed.
• Use  and  to toggle between Serial and GPIB interface when
the Interface menu is displayed.
• Use  and  to scroll from Address:1 to Address 30 when the
address menu is displayed.
Selects Output mode.
Selects Input mode.
Zeros the pressure module reading when in Pressure Measurement
mode.
Zeros the thermocouple TC mV/°C offset when in TC Measurement
mode.
Toggles between Centigrade and Fahrenheit when you are using
the TC or RTD functions.
AUTOSET runs through the setpoints you entered using the SET
function. Press  . AUTO SET POINT? appears on the
display.
Enter a number between 1 and 9 that corresponds to the number of
setpoints being used. DWELL TIME 5-500? appears on the display.
Dwell time is the number of seconds between each setpoint. The
output cycles through each setpoint and then reverses the order.
For example, if 5 is entered for the number of setpoints, the
Calibrator cycles through setpoints 1, 2, 3, 4, 5 and then reverses to
setpoints 4, 3, 2, and 1.
Setpoints of 30 V or greater will not go to standby
when you use this feature.
Enters an exponent when you define a custom RTD.
Getting Acquainted with the Calibrator
Description
Caution
Getting Started
1