General
Safety information and protective measures
CAUTION
Siemens-Elema AB
Solna, Sweden
•
When performing service work and tests adhere to:
- the product-specific safety information in the documents,
- as well as the general safety information contained in ARTD
Part 2.
•
Connect the MOBILETT Plus HP only to mains power supply
outlet (receptacle), that corresponds to the installation
requirements of VDE 0107 or country-specific regulations.
•
Remove or install boards only when the generator is switched
OFF. Adhere to the ESD guidelines
•
Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol
types of adjustments, radiation protective clothing must be
worn.
SPR8-220.033.03
Rev. 02
.
Page 3 of 8
10.02
SPS-UD
1 - 3
1
. During these
MOBILETT Plus HP