Fluke 8846A Programmer's Manual page 59

Hide thumbs Also See for 8846A:
Table of Contents

Advertisement

CALCulate:FUNCtion(?) {NULL|DB|DBM|AVERage|LIMit}
Description Selects one of five math functions. This setting is stored in volatile
memory. Selecting a math function disables any math function which may
have been set previously. For the selected math function to take effect, the
CALCulate:STATe must be ON.
Parameters <n>
Examples
CALC:FUNC DBM
Query
CALC:FUNC?
CALCulate:STATe(?) <n>
Description Activates or deactivates the math function selected with CALC:FUNC
command. This setting is stored in volatile memory.
Parameters <n>
Examples
CALC:STAT ON
Query
CALC:STAT?
CALCulate:AVERage:MINimum?
Description Retrieves the minimum value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or receives a reset through the remote
interface. This setting is stored in volatile memory.
Query CALC:AVER:MIN?
CALCulate:AVERage:MAXimum?
Description Retrieves the maximum value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or receives a reset through the remote
interface. This setting is stored in volatile memory.
Query CALC:AVER:MAX?
CALCulate:AVERage:AVERage?
Description Retrieves the average value measured during a statistics operation. This
value is cleared whenever the Meter is powered off then back on, a
statistics function is enabled, or receives a reset through the remote
interface. This setting is stored in volatile memory.
Query CALC:AVER:AVER? Returns the maximum value of statistics operation.
=
NULL
Uses the present reading as an offset to zero
the returned reading.
DB
Selects the DB mode.
DBM
Selects the dBm mode.
AVERage
Selects the average mode
LIMit
Selects the Limit Testing mode.
Selects the dBm mode.
Returns the selected math function.
=
ON or 1
Activates the math function set previously by
CALC:FUNC command.
OFF or 0
Deactivates the math function.
Turned on the math function.
Returns the state of the math function. (0 = OFF
and 1 = ON)
Returns the minimum value of statistics operation.
Returns the maximum value of statistics operation.
Digital Multimeter
Supported SCPI Commands
51