Questionable Data
Event Register
Enable Register
0
Voltage Overload
1
Current Overload
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
9
Ohms Overload
Not Used
11
Limit Test Fail LO
12
Limit Test Fail HI
13
Remote Mode
Not Used
15
Not Used
STAT:QUES:EVEN?
STAT:QUES:ENAB <value>
STAT:QUES:ENAB?
Standard Event
Event Register
Enable Register
0
Operation Complete
Not Used
2
Query Error
3
Device Error
4
Execution Error
5
Command Error
Not Used
7
Power On
*ESR?
*ESE <value>
*ESE?

Standard Event Status and Standard Event Status Enable Registers

The Standard Event Status Register (ESR) assigns events to specific bits (see Figure 3
and Table 5). When a bit in the ESR is set (for example, 1), the event that corresponds to
that bit has occurred since the register was last read or cleared. For example, if bit 3
(DDE) is set to 1, a device-dependent error has occurred.
The Standard Event Status Enable Register (ESE) is a mask register that allows the host
to enable or disable (mask) each bit in the ESR. When a bit in the ESE is 1, the
corresponding bit in the ESR is enabled. When any enabled bit in the ESR changes from
O to 1, the Standard Event bit in the Status Byte Register also goes to 1. When the ESR is
read (using the *ESR? command) or cleared (using the *CLS command), the Standard
Event bit in the Status Byte Register returns to 0.
"OR"
"OR"
Figure 3. Overview of Status Data Structure
Service Requests (IEEE 488 Only) and Status Registers
Binary Wieghts
0
8
2
=
1
2
=
256
1
9
2
=
2
2
=
512
2
10
2
=
4
2
= 1024
3
11
2
=
8
2
= 2048
4
12
2
= 16
2
= 4096
5
13
2
= 32
2
= 8192
6
14
2
= 64
2
= 16384
7
15
2
= 128
2
= 32768
Status Byte
Summary Register Enable Register
0
Not Used
Not Used
Not Used
3
Questionable Data
4
Message Available
5
Standard Event
6
Request Service
7
Not Used
Serial Poll (SPOLL)
*SRE <value>
*STB?
*SRE?
Output Buffer
Digital Multimeter
"OR"
caw01f.eps