Mild System Contamination Cleanup; Procedure - Welbilt Manitowoc NEO UDE0065A Technician's Handbook

Undercounter ice machines
Hide thumbs Also See for Manitowoc NEO UDE0065A:
Table of Contents

Advertisement

MILD SYSTEM CONTAMINATION CLEANUP

PROCEDURE

1. Replace any failed components.
2. If the compressor is good, change the oil.
3. Replace the liquid line drier.
NOTE: If the contamination is from moisture, use heat
lamps during evacuation. Position them at the compressor,
condenser and evaporator prior to evacuation. Do not
position heat lamps too close to plastic components, or
they may melt or warp.
4. Follow the normal evacuation procedure, except
replace the evacuation step with the following:
A.
Pull vacuum to 1000 microns. Break the vacuum
with dry nitrogen and sweep the system.
Pressurize to a minimum of 5 psig.
B.
Pull vacuum to 500 microns. Break the vacuum
with dry nitrogen and sweep the system.
Pressurize to a minimum of 5 psig.
C.
Change the vacuum pump oil.
D.
Pull vacuum to 500 microns. Run the vacuum
pump for 1/2 hour on self-contained models, 1
hour on remotes.
NOTE: You may perform a pressure test as a preliminary
leak check. You should use an electronic leak detector after
system charging to be sure there are no leaks.
5. Charge the system with the proper refrigerant to the
nameplate charge.
6. Operate the ice machine.
212
Part Number 000014797 Rev02 05/19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents