Using the calibration
offset function
MEASURED TEMP.
HEATER
ALARM
AUTO STOP
AUTO START
SET
TEMP.
FIXED TEMP.
PROGRAM
OVER TEMP .
PROTECTOR
RUN
ENTER
STOP
FIXED
SUB
TIMER
PROGRAM
TEMP.
MENU
4. How to operate the unit
Useful functions(calibration offset function)
The calibration offset function compensates any difference between
the target in-bath temperature and the control temperature of the
controller (sensor temperature). You can apply parallel compensation
to the plus or minus side over the entire temperature range of the unit.
You can set/cancel this function with the Sub menu key.
① Start operation at the target set temperature and check the in-bath
temperature (sample temperature) on a temperature recorder
when the temperature is stable.
② Check the difference between the set temperature and the in-bath
temperature (sample temperature).
③ Press the Sub menu key again to select the character
cAL
that mean calibration offset and then press the
Start/Stop key.
④ Enter the difference between the set temperature and the in-bath
▽ and △
temperature with
longer to finish setting.
※ You can set an offset compensation temperature to either + or –
side up to +99℃ and -99℃ respectively..
Setting to the - side will decrease the temperature on the
measured temp indicator by the compensated temperature and
the in-bath temperature will increase by that amount accordingly.
Setting to the + side will increase the temperature on the
measured temp indicator by the compensated temperature and
the in-bath temperature will decrease by that amount
accordingly.
※ Inputting a too large compensation may make difference
between the actual temperature and the displayed temperature
larger and cause a dangerous situation. Contact a nearest sales
office or our general customer service center before inputting a
larger compensation value.
※ The unit has a two-point compensation function that adjusts
offset for the lower temperature region and the high temperature
region in addition to the calibration offset function and adjusting
temperatures have been set at the time of factory shipping.
※ When validating the temperature indicator, first consult with your
nearest sales office or the customer support center.
34
Controlled temp after
compensation to the
minus side
Present temperature
Controlled temp after
compensation to the
plus side
keys and press the Sub menu key