Drystar 5301 XL, Drystar 5302 XL | Advanced Operation (key-operator mode) | 65
Density response of film
Different densitometers can give different results when measuring the same
density area. This is due to spectral differences between the densitometers.
The density values used in this manual are based on the X_Rite 310
densitometer. Values measured with another densitometer can vary from
these values.
It's possible to change the default densitometer. You can also compare the
density values from different densitometers with each other.
Non-calibrated or badly calibrated densitometers can result in even larger
differences.
The user has a selection of common densitometers. All shown densities will
vary slightly according to the densitometer selection.
Related Links
Film density response data
on page 132
Changing general image quality settings - Default densitometer
on page 38