Order-No.: DD+DIS015.02E
Work Instructions for order no.
The maintenance has to be carried out:
·
Once a year or every 25 000 cycles
Maintenance must be carried out according to the maintenance instructions,
Maintenance Step by Step
Diagnostics
Inside
Cassette Unit
Power Unit
cPCI-Rack
Scan Unit
Transport Units
printed in Germany 01 / 2002
see chapter 9 of your Service Documentation.
·
Ask customer about problems
·
Read out the infocounters, analyse them and clear
them afterwards
·
Check visually the overall condition
·
Vacuum the inside and wipe it
·
Clean opener mechanism
·
Check roller for visible wear and exchange if necessary
·
Exchange belt once a year
·
Check electrical function of safety switch
·
Exchange air filter
·
Remove dust and dirt
·
Check electrical function of the cPCI-Rack fan
·
Clean grid of the fan and exchange air filter
·
Clean scan-rollers with ADC Cleaner
(do not remove!)
·
Check discharge brush and clean or replace it if
necessary
·
Check suction cups position, clean them with ADC
Cleaner and exchange suction cups once a year