AGFA ADC Compact Plus Maintenance Manual page 13

Hide thumbs Also See for ADC Compact Plus:
Table of Contents

Advertisement

Maintenance
Lamps
Chapter 9 / 10
(3) Remove the complete
Erasure Unit
(4) Remove the pane 4
(5) Remove all lamps 1
(6) Dust the following parts:
- Reflector
- Input and output opening of
the air stream (protection
grid) 2
- KG2 filter 3
- Outer front panel
(7) Check the KG2 filter 3 for
damage and replace if
necessary.
(8) Insert new lamps every
maintenance, for this use a
dry cloth!
Do not touch the lamps
with bare fingers! Use a
soft cloth to insert the
lamp.
(9) Re-insert the Erasure Unit.
ADC Compact Plus
(Type 5146 / 100/200)
In case of persistent dirt, you may also use
ADC Cleaner for cleaning all the surfaces
with the exception of the inner side of the
large glass plate. This side must not be
cleaned with anything wet since a gelatin
layer is attached to it.
1
4
Figure 10
DD+DIS015.02E
2
3
Edition 1, Revision 1