12.1. Inspection and safety management

Do not open the device. It uses delicate electronic components and an intricate air unit that
could be damaged. If you cannot fix the problem using the troubleshooting instructions, request
service from your local dealer or from the A&D service group. The A&D service group will
provide technical information, spare parts and units to authorized dealers.
Technical inspection procedures which should be done at least every two years, can be
performed either by the manufacturer or by an authorized repair service in accordance with the
regulations governing manufacturing of medical products.
Checking pressure accuracy
When using a rubber pump, do not apply a pressure of 280 mmHg or higher to
the monitor or inspection equipment (UM-102, accurate mercury
sphygmomanometer or aneroid gauge).
Perform the inspection only as described below or the setting values and
function settings may be changed.
After inspection, check that the air connector plug is inserted into the blood
pressure monitor. If the air connector plug is not inserted, pressure cannot be
applied and measurement is not possible. When inserting the plug, push in until
you hear a click.
Objective:
Inspection equipment: Inspection equipment (UM-102, Accurate mercury sphygmomanometer
Connection:

MAINTENANCE

12.
Compare the pressure values of the inspection equipment and the blood
pressure monitor to check for errors in the monitor.
or aneroid gauge)
Connect the inspection equipment to the blood pressure monitor as
shown below. Remove the armrest of the blood pressure monitor and
then remove the cover of pressure inspection area. Remove the air
connector plug from the air socket of the blood pressure monitor.
Connect the coupling connector to the connection hose, and connect it
to the air socket.
Caution
41