A&D UA-651BLE Instruction Manual

A&D UA-651BLE Instruction Manual

Digital blood pressure monitor
Hide thumbs Also See for UA-651BLE:

Advertisement

Digital Blood Pressure Monitor
Model UA-651BLE
Instruction Manual
Original
'
Manuel d
instructions
Traduction
Manual de Instrucciones
Traducción
Manuale di Istruzioni
Traduzione
Bedienungsanleitung
Übersetzung
使用手冊
翻譯
1WMPD4003018

Advertisement

Table of Contents
loading

Summary of Contents for A&D UA-651BLE

  • Page 1 Digital Blood Pressure Monitor Model UA-651BLE Instruction Manual Original ’ Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione Bedienungsanleitung Übersetzung 使用手冊 翻譯 1WMPD4003018...
  • Page 2: Table Of Contents

    Contents Dear Customers ...............2 Preliminary Remarks..............2 Precautions ................2 Parts Identification ..............4 Symbols ...................5 Using the Monitor..............7 Installing / Changing the Batteries ........7 Connecting the Air Hose ............7 Connecting the AC Adapter..........7 Wireless Function..............8 Bluetooth® Transmission .............8 Pairing .................9 Selecting the Correct Cuff Size ...........11 Applying the Arm Cuff ............12 How to Take Accurate Measurements ........12 Measurement..............12...
  • Page 3: Dear Customers

    (0123: The reference number to the involved notified body) Hereby, A&D Company, Limited declares that the radio equipment type UA-651BLE is in compliance with Directive 1999/5/EC. The full text of the EU declaration is available at the following internet address: http://www.aandd.jp/products/manual/manual_medical.html...
  • Page 4 Wireless communication devices, such as home networking devices, mobile phones, cordless phones and their base stations, walkie-talkies can affect this blood pressure monitor. Therefore, a minimum distance of 3.3 meters should be kept from such devices. When reusing the device, confirm that the device is clean. Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regulations.
  • Page 5: Parts Identification

    Parts Identification Proper Fit Range Index Artery Position Mark Air Hose DC Jack Arm Cuff AC Adapter Plug Display START Button Air Socket Air Connector Plug Battery Compartment Battery Cover 1.5V Batteries (LR6, R6P or AA) Display Systolic Pressure Diastolic Pressure Pressure Bar Indicator Pulse Rate Heart Mark...
  • Page 6: Symbols

    Symbols Symbols that are printed on the device case Symbols Function / Meaning Standby and Turn the device on. Systolic blood pressure in mmHg SYS. Diastolic blood pressure in mmHg DIA. Pulse per minute PUL. Battery installation guide Direct current Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks.
  • Page 7 Symbols Function / Meaning Recommended Action The battery power indicator during measurement. Full Battery Replace all batteries with new ones when The battery is low when it blinks. Low Battery the mark blinks. Take another measurement. Unstable blood pressure due to movement during measurement.
  • Page 8: Using The Monitor

    Using the Monitor Step 2 Installing / Changing the Batteries Remove the battery cover. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that Step 1 the polarities (+) and (-) are correct. Step 2 Step 3 Replace the battery cover.
  • Page 9: Wireless Function

    Using the Monitor Wireless Function CAUTION In the unlikely event that this monitor causes radio wave interference to a different wireless station, change the location where this monitor is used or stop use immediately. Be sure to use in a location where visibility between the two devices that you want to connect is good.
  • Page 10: Pairing

    Using the Monitor Pairing A Bluetooth® device needs to be paired with a different specific device in order to communicate with that device. If this monitor is paired with a receiver device from the start, measurement data is transmitted automatically to the receiver device each time a measurement is made.
  • Page 11 Using the Monitor Communication distance The communication distance between this monitor and the receiver device is about 10 m. This distance is reduced by the conditions in the surrounding environment, so be sure to check that the distance is short enough for a connection to be made after measurement is complete.
  • Page 12: Selecting The Correct Cuff Size

    CUF-F-LA 22 cm to 32 cm Adult cuff CUF-F-A Arm size: The circumference of the biceps. Note: Model UA-651BLE is not designed for using a small adult cuff. Symbols that are printed on the cuff Symbols Function/Meaning Recommended Action Set the...
  • Page 13: Applying The Arm Cuff

    Using the Monitor Applying the Arm Cuff Do not roll up shirtsleeve tightly 1. Wrap the cuff around the upper arm, about 1-2 cm above the inside of Cuff the elbow, as shown. 1 to 2 cm Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error.
  • Page 14: Measurements

    Press the button again to turn START Pulse rate off the power. Exhausts remaining air automatically Note: Model UA-651BLE is provided with an automatic power shut-off function. Allow at least three minutes between measurements on the same person. English 13...
  • Page 15: Measurement With The Desired Systolic Pressure

    Measurements Measurement with the Desired Systolic Pressure Model UA-651BLE is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If re-inflation occurs repeatedly or your systolic blood pressure is expected to exceed 230 mmHg, use the following method.
  • Page 16: What Is An Irregular Heartbeat

    What is an Irregular Heartbeat The UA-651BLE blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement.
  • Page 17: About Blood Pressure

    About Blood Pressure What is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
  • Page 18: Troubleshooting

    described in this manual to get to know your Typical fluctuation within a day (Measured every five minutes) normal blood pressure. Regular readings give Systolic a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure.
  • Page 19: Maintenance

    However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area or A&D for maintenance. Technical Data Type UA-651BLE Measurement method Oscillometric measurement Measurement range Pressure: 0 - 299 mmHg...
  • Page 20 Useful life Device: 5 years (when used six times a day) Cuff: 2 years (when used six times a day) Ingress protection Device: IP20 Accessory AC adapter The adapter is to connect the blood pressure monitor to a power source at home. TB-233 Please contact your local A&D dealer for purchasing.
  • Page 21 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A.