BLOOD PRESSURE MEASUREMENT

7.
To stop blood pressure measurement halfway through, press the START/STOP
button. The cuff rapidly deflates and returns to its original state.
If measurement cannot be stopped by pressing the START/STOP button,
press the FAST STOP button (on the front of the monitor).
1. Insert bare arm or arm with a thin
shirt into the arm insertion section
up to the top of the shou+lder.
(If thick clothing is worn, the
measurement results will be
incorrect. Remove thick clothing
before measurement.)
2. Press the START/STOP button to
start blood pressure measurement.
3. The cuff automatically inflates.
Keep the arm still in the cuff during the
measurement.
4. After inflation, deflation starts automatically. As the
pressure decreases, measurement is performed.
The patient must relax and remain still. (See "10.3.
Applied pressure")
5. After about one minute of measurement, the cuff
automatically deflates to its original state.
6. The measurement results are displayed.
7. The measurement results are printed on the printer paper. Remove the arm from the cuff.
(See "10.5. Print quality")
When performing continuous measurements, wait 2 to 3 minutes between measurements for the
patient to relax.
Blood pressure measurement results are affected by the posture and physical condition of the patient.
If the patient moves or talks during measurement, correct measurement is not possible.
To obtain accurate measurement results, ensure the patient sits with good posture and his/her
back straight, and with his/her feet flat on the floor without crossing legs. Ensure the patient is
relaxed and remains still.
Adjust the height of the chair such that the cuff is at the same height as the heart. If the cuff is
not at the same height as the heart, correct measurement is not possible.