Maintenance

Do not open the device. It uses delicate electrical components and
an intricate air unit that could be damaged. If you cannot fix the
problem using the troubleshooting instructions, contact the authorized
dealer in your area or our customer service department. A&D customer
service can provide technical assistance and spare parts.
Technical Data
Type
Measurement method
Measurement range
Measurement accuracy
Power supply
Rating
Number of measurements
Classification
Clinical test
Wireless communication
E-25
UA-651BLE
Oscillometric measurement
Pressure: 0-299 mmHg
Systolic pressure: 60-279 mmHg
Diastolic pressure: 40-200 mmHg
Pulse: 40-180 beats /min
Pressure: ±3 mmHg
Pulse: ±5 %
4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included)
DC6V 3W
Approx. 700 times LR6 (alkaline
batteries) Approx. 200 times R6P
(manganese batteries)
With pressure value 180 mmHg, room
temperature 23 °C.
Internally powered ME equipment
(Supplied by batteries)/Class II
(Supplied by adapter) Continuous
operation mode
According to ISO81060-2:2013
In the clinical validation study, K5 was
used on 85 subjects for determination of
diastolic blood pressure.
Bluetooth Ver. 5.1LE BLP