Errors And Troubleshooting - Braun ThermoScan 7+connect Manual

Ear thermometer
Table of Contents

Advertisement

Available languages

Available languages

Errors and troubleshooting

Error message
Situation
No lens filter is attached.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(50 – 104 °F or 10 – 40 °C).
Temperature taken is not within
typical human temperature range
(93.2 – 108 °F or 34 – 42.2 °C).
HI = too high
LO = too low
System error – self-check display
flashes continuously and will not be
followed by the ready beep and the
ready symbol or screen is blank.
If error persists,
If error persists,
If error still persists,
Unsucessful connection.
10
Solution
Attach new, clean lens filter.
Take care that the positioning of the
probe is correct and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new
measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between
50 and 104 °F or 10 and 40 °C.
Make sure the probe tip and lens are
clean and a new, clean lens filter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
... reset the thermometer by removing
the batteries and putting them back in.
... replace the batteries with new
batteries.
... please contact the Service Center.
Check issue in the app for more details.
Errors and troubleshooting
Battery is low, but thermometer will
still operate correctly.
Battery is too low to take correct
temperature measurement.
FCC warning
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
A. Reorient or relocate the receiving antenna.
B. Increase the separation between the equipment and receiver.
C. Connect the equipment into an outlet on a circuit dierent from that to which the receiver is
connected.
D. Consult the dealer or an experienced radio/TV technician for help.
Any changes or modications not expressly approved by the party responsible for compliance
can void the user's authority to operate the equipment
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Insert new batteries.
Insert new batteries.
11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6575

Table of Contents