Preliminary Adjustment of the Well-plate Handler Position

These adjustments are made by moving the Well-plate Handler arm and
gripper fingers manually. These actions do not damage the Well-plate Handler
mechanisms.
1 Turn off the power on the back panel of the Well-plate Handler.
2 Loosen the six adjusting screws holding the Well-plate Handler in position.
3 Place a plate in the rack R1.
4 Position the gripper arm above rack R1.
5 Move the arm down until it touches the plate.
6 Squeeze the gripper fingers so that they grip the plate.
7 Keeping the gripper fingers closed, raise the arm until the plate is clear of
8 Rotate the arm until it is positioned above the Automation Interface.
9 Lower the arm until the rack is placed on the transport position of the
10 Adjust the position of the Well-plate Handler until the plate is squarely and
11 Tighten the adjustment screws.
If you have fitted the optional rack extension, you must also adjust its lateral
position.
1 Loosen the three adjusting screws holding the rack extension in position.
2 Place a plate in the rack R1.
3 Position the gripper arm above rack R1.
4 Move the arm down until it touches the plate.
5 Squeeze the gripper fingers so that they grip the plate.
6 Keeping the gripper fingers closed, raise the arm until the plate is clear of
7 Rotate the arm until it is positioned above rack R5.
8 Lower the arm until the rack is immediately above the rack
1100 Series Sample Capacity Extension
the rack.
Automation Interface.
centrally positioned on the transport position of the Automation Interface.