Solid
0.14 to 2.5 mm
cable
wire
Stranded
0.2 to 2.5 mm
wire
AWG
24 to 12
Wire
12 mm
stripping
length
Degree of protection
IP20
Housing material
PBT
Weight
54 g
Safety standards
UL 508
EMC
EMI:
Radiated EMI:
EMS:
ESD immunity:
Rated electromagnetic field immunity:
Burst immunity:
Surge immunity:
Conducted disturbance immunity:
Note: K is the abbreviation for Kelvin, the unit for absolute
temperature.
Nomenclature
18
Thermocouple Transducer
2
2
EN 55011
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
Diagnostics indicator (red)
DIP switch S3
5 6 Output signals
7 8 Supply voltage
K3FP-TS-UI
■ Input Specifications
Input signal
Item
Measurement range J: − 150 to 1,200 ° C
Minimum
measurement span
(See note.)
Max. input signal
Note: Set the start and end temperatures using the DIP switch to
50 ° C or higher.
■ Output Specifications
Output signal
Item
Allowable load
impedance
Max. output signal
Non-load voltage
Short-circuit current ---
Ripple
Operation during
sensor faults
Translucent cover
2 3 Inputs: Type J or K thermocouple sensor
DIP switch S2
DIP switch S1
Snap for mounting DIN rail
DIN rail bus connection
Type J and K Thermocouples
(Conforms to IEC 60584-1)
K: − 150 to 1,350 ° C
Min. 50 ° C
30 V
0 to 20 mA DC,
0 to 5 V DC,
4 to 20 mA DC,
1 to 5 V DC,
20 to 0 mA DC,
0 to 10 V DC,
20 to 4 mA DC
10 to 0 V DC
500 Ω max.
10 k Ω min.
23 mA
12.5 V
12.5 V max.
---
10 mA
20 mV pp max.
20 mV pp max.
(500 Ω )
0% to 105%