Troubleshooting; Specifications - HoMedics THERA P BPW-051DDM Manual

Automatic wrist blood pressure monitor
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING

If any abnormality arises during use, please check the following points:
SYMPTOMS
Unit does not turn on
Batteries have
when the
run down.
START/STOP
button is pushed.
Battery polarities have
been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
blood pressure value
Did you talk or move
is displayed exces-
during
sively low
measurement?
(or high).
Shaking of the wrist
with the cuff on.
E1 error symbol
Air circuit
shown on display
abnormality.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Data error.
shown on display
"
--" appears on
Recalibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page (See page 35).
POSSIBLE
CORRECTION
CAUSES
Replace them
with two new AAA
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again.
Keep wrist steady
during measurement.
Measure again.
Switch the unit off,
then measure again.
Remove and reload the
batteries.
Refer to page 28 for
more information.
33

SPECIFICATIONS

Power Source:
Two 1.5V DC (AAA) Alkaline batteries
Measurement
Oscillometric
Method:
Pressure: 40~250 millimeters Mercury
Measurement
Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity: 60 memories for each User (120 total)
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 50°F~104°F (10°C~40°C)
Environment:
Humidity: 40%~85% RH max.
Storage
Temperature: 14°F~140°F (-10°C~60°C)
Environment:
Humidity: 10%~90% RH max.
Weight:
0.25 lbs (without batteries)
Wrist
5.3"- 8.6" (13.5 - 22cm)
Circumference:
3.25"(L) x 2.75"(W) x 2.60"(H)
Dimensions:
82mm(L) x 69mm(W) x 66mm(H)
Type BF
Device and cuff are designed to
provide special protection against
electrical shocks.
Note: These specifications are subject to change
without notice.
(mmHg)
34

Advertisement

Chapters

Table of Contents
loading

Table of Contents