J MAXIMUM SENSITIVITY SETTING
(TEACHING)
Set the mode selector to "SET" to start teaching. The red level
display will flash if teaching error occurs. In this case, repeat the
entire teaching procedure.
Maximum Sensitivity Setting
1. Set the "SET ADJ RUN" switch to "SET" position.
2. Press the "TEACH" button for 3 seconds (min).
3. Teaching is complete when the level display changes from
red to green. Digital incident level will appear shortly after the
display changes from red to green.
(Red)
(Green)
1234
4. Set the "SET ADJ RUN" switch to "RUN" position to imple-
ment settings from the teach function.
8
J ONE-POINT TEACH MODE
(WITHOUT OBJECT)
1. Set the "SET ADJ RUN" switch to "SET" position.
2. Press the "TEACH" button for approximately 1 second.
3. Teaching is complete when the red level display is lit.
Digital incident level will appear shortly after the red level
display is lit.
(Red)
1234
4. Set the "SET ADJ RUN" switch to "RUN" position to imple-
ment settings from the teach function.
5. The threshold is automatically set.
Note: Try the two-point teaching mode if the difference in level is
too fine.